Assemble the actuators
There is one "actuator column" for each of the three axes of the OpenFlexure Microscope stage. These allow you to move the sample in X and Y, or focus the microscope by moving in Z.
For this section you will need:
Printed Parts
- 4 feet - Each actuator has its own labelled foot.
- 4 large gears
Printed Tools
- 1 band tool
- 1 band tool cover
- 1 nut tool
Sub-Assemblies
Consumables
- 4 drops of light oil - Don't skip this or you will damage the screws
Mechanical Components
- 4 M3 brass nut
- 8 M3 stainless steel washers
- 4 M3x25mm stainless steel hex bolt
- 4 Viton O-ring (30mmx2mm) - "Viton band"
Step 1: Insert the nut
- Insert a brass nut into the x actuator through the hole in the front of the main body.
- Looking through the hole you should see the side of the nut. If it is tilted so you can see the top can tap the microscope until it sits flat.
Step 2: Attach the gear
- Line up two washers so that the slightly curved sides are facing each other.
- Take one of the hex bolts
- Push it through one of the large gears and then the two washers.
- Push the screw through the hole at the top of the x actuator until it reaches the nut
- Screw the bolt into the nut until the nut is completely lifted up. (If the nut turns hold it in place with the nut tool)
Step 3: Oil the lead screw
- Lift the gear so you can see about 5mm of screw thread underneath
- Add one drop of light oil onto the screw thread
Step 4: Assemble the band insertion tool
- Insert the two arms of the band tool into the band tool cover.
Step 5: Prepare the actuator
- Look through the hole you inserted the nut into
- You should be able to see the screw thread of the hex bolt
- Rotate the gear until the screw is clearly visible through this hole
- Push the nut tool into the hole to hold the internal actuator in place
Step 6: Attaching the viton bands and foot
This is the trickiest part of the microscope build. In this step we will clip bands onto hooks inside each actuator.
- Take the foot for the x actuator
- Loop a viton band through the foot
- Push the assembled band tool through the foot and hook the band onto the tool on each side.
- Align the foot under the microscope so that the letter faces outwards
- Check that the nut tool is still blocking the actuator column.
- Push the band tool into the microscope until it clicks
- Remove the band tool.
If you had problems with this step see our troubleshooting page.
Step 7: Seat the foot
- If the foot did not click into place during actuator assembly align it and push it into place
- The foot should sit flush with the base of the microscope
- The front of the foot should align with the front of the actuator
Step 8: Repeat this process for Y and Z
Follow the same procedure for the Y and Z axes. Once complete the microscope with assembled actuators should look like this:
Step 9: Repeat this process for separate z-actuator
Follow the same procedure for the separate z-actuator. Once complete the complete separate z-actuator should look like this: